Description
USED EQUIPMENT without the Option SC300
The equipment is in very good cosmetic and work conditions.
Technical Specifications
Voltage DC
Range: 0 to ±1020V
Best 1 Year Specification: ±50 ppm of setting
Current DC
Range: 0 to ±11A
Best 1 Year Specification: ±0.01% of setting
Resistance
Range: 0 to 3.29.999 MΩ
Best 1 Year Specification: ±0.009% of setting
Voltage AC
Range: 1 mV to 1020V
10 Hz to 500 kHz, Sine
Best 1 Year Specification: ±0.03% of setting
Current AC
Range: 29 m A to 11A
10 Hz to 10 kHz, Sine
Best 1 Year Specification: ±0.06% of setting
Capacitance
Range: 0.33 nF to 1.1 mF
Best 1 Year Specification: ±0.25% of setting
Thermocouple Source and Measure
Range: -250°C to +2316°C (11 types)
Best 1 Year Specification: ± 0.14°C of setting
RTD Source
Range: -200°C to +630°C
(8 types)
Best 1 Year Specification: ±0.03°C of setting
DC Power
Range: 109 mW to 11 kW
Best 1 Year Specification: ± 0.08% of setting
AC Power
Range: 109 mW to 11 kW
45-65 Hz, PF=1
Best 1 Year Specification: ± 0.15% of setting
Phase
Range: 0 to ± 179.99°
Best 1 Year Specification: ±0.15°
Frequency
Range: 0.01 Hz to 2.0 MHz
Best 1 Year Specification: ±25 ppm of setting
General Specifications
Warm up Time
Twice the time since last warmed up, to a maximum of 30 minutes
Settling Time
Less than 5 seconds for all functions and ranges except as noted
Standard Interfaces
IEEE-488 (GPIB), RS-232, 5725A (5500A only)
Temperature Performance
Operating: 0°C to 50°C
Calibration: (tcal): 15°C to 35°C
Storage: -20°C to 70°C
Temperature Coefficient
Temperature coefficient for temperatures outside tcal ± 5°C is 10% of the 90-day specification (or 1-year, as applicable) per °C
Relative Humidity
Operating: < 80% to 30°C, < 70% to 40°C, < 40% to 50°C
Storage: < 95%, non-condensing
Altitude
Operating: 3,050m (10,000 ft) maximum
Non-operating: 12,200m (40,000 ft) maximum
Safety
Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92
Analog Low Isolation
20V
EMC
Designed to comply with FCC Rules Part 15
Line Power
Line Voltage (selectable): 100V, 120V, 220V, 240V
Line Frequency: 47 to 63 Hz
Line Voltage Variation: : ± 10% about line voltage setting
Power Consumption
300 VA
Calibration Documentation 17025 accredited report of calibration included
Dimensions
Height: 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit
Width: 43.2 cm (17 in), standard rack width
Depth: 47.3 cm (18.6 in) overall
Weight (without options)
20 kg (44 lbs.)
Absolute Uncertainty Definition
5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated.
Specification Confidence Interval
> 99%